BHD Rentals 800.565.3317 info@bhdrentals.com

Hart 1524-P1-156 1524 Digital Thermometer Kit - 2 Channel Data Logger

Overview

The 1524 Reference Thermometers from Fluke Calibration measure, graph, and record PRTs, thermocouples, and thermistors. These thermometer readouts deliver exceptional accuracy, wide measurement range, temperature logging, and trending, all in a handheld tool you can take anywhere.

The 1524 Reference Thermometers from Fluke Calibration measure, graph, and record PRTs, thermocouples, and thermistors. These thermometer readouts deliver exceptional accuracy, wide measurement range, temperature logging, and trending, all in a handheld tool you can take anywhere.

The 1524 lets you handle field applications, laboratory measurements, and data logging with ease. And with the dual channel measurement capabilities of the model 1524, you can do twice the work in half the time.

Two channels; memory for logging 15,000 measurements; real-time clock for time and date stamps

Three sensor types

  • PRTs: -200 °C to 1000 °C
  • Thermocouples -200 °C to 2315 °C
  • Precision thermistors: -50 °C to 150 °C

 

Accuracy

  • PRTs: up to ±0.011 °C
  • Thermocouples : ±0.24 °C for J,K,L,M
  • Precision thermistors: ±0.002 °C

 

Fast mode

  • PRTs: 0.45 seconds per sample
  • Thermocouples: 0.3 second per sample
  • Precision thermistors: 0.3 seconds per sample

 

Graphical display

  • 128x64 backlit LCD graphic display
  • Plot and scale trends in real time
  • Simultaneous dual channel readings

 

Specifications
Input channels 2
Logging 15,000 time and date stamped; 25 readings with statistics
Sample interval
(normal)
1 seconds
Sample interval (fast mode) 0.3 seconds (see technical manual for details)
Sensor types PRTs, RTDs, thermistors, and thermocouples
Thermocouple types C,E,J,K,L,M,N,T,U,B,R,S
Operating temperature -10 C to 60 C (Best accuracy 13 C to 33 C)
Power requirements 3 AA alkaline batteries
Size 96 x 200 x 47 mm
(3.75 x 7.9 x 1.86 inches)
Weight  0.65 kg (1.4 lb)
Environmental conditions for best accuracy  13 C to 33 C
Millivolt range and accuracy 10 mV to 75 mV (0.005 % + 5 μV)
Resistance range and accuracy 0 Ω to 400 Ω (0.004 % + 0.002 Ω)
200 Ω to 50 kΩ (0.01 % + 0.5 Ω) 
50 kΩ to 500 kΩ (0.03 %)
Temperature coefficient, voltage ( 10 C to 13 C , +33 C to 60 C) (0.001 %/C + 1 μV/C)
Temperature coefficient, resistance ( 10 C to 13 C , +33 C to 60 C) 0.0008 %/C + 0.0004 Ω(0 Ω to 400 Ω)
0.002 %/C + 0.1 Ω(0 Ω to 50 kΩ)
0.06 %/C + 0.1 Ω(50 kΩ to 500 kΩ)
Excitation current, resistance 1 mA (0 Ω to 400Ω)
10 μA (0 Ω to 50 kΩ)
2 A (50 kΩ to 500 kΩ)
Thermocouple Equivalent Temperature Accuracies (Readout Only)
Type B 0.85 C from 600 C to 800 C
0.68 C from 800 C to 1000 C
0.57 C from 1000 C to 1800 C
Type C 0.32 C from 100C to 550 C
0.71 C from 550 C to 2300 C
Type E 0.52 C from -200 C to 0 C
0.22 C from 0 C to 950 C
Type J 0.52 C from -200 C to 0 C
0.23 C from 0 C to 1200 C
Type K   0.61 C from -200 C to 0 C  
0.24 C from 0 C to 1370 C
Type L 0.36 C from -200 C to 0 C
0.23 C from 0 C to 1370 C
Type M 0.26 C from -20 C to 0 C
0.25 C from 0 C to 400 C
0.22 C from 400 C to 1400 C
Type N 0.72 C from -200 C to 0 C  
0.28 C from 0 C to 1300 C
Type R 1.09 C from -20 C to 0 C
0.97 C from 0 C to 500 C 
0.49 C from 500 C to 1750 C
Type S 1.05 C from -20 C to 0 C
0.95 C from 0 C to 500 C 
0.56 C from 500 C to 1750 C
Type T 0.60 C from -200 C to 0 C
0.25 C from 0 C to 400 C
Type U 0.54 C from -200 C to 0 C
0.24 C from 0 C to 400 C
Note 1: Accuracies are based on internal reference junction compensation. refer to technical manual for equivalent accuracies with an external reference junction.
1523/24 Accuracies with Selected Probes (C)
-200 C 5616-12: 0.014
5615-6: 0.025
5627A-12: 0.027
5610-9: n/a
0 C 5616-12: 0.021
5615-6: 0.021
5627A-12: 0.049
5610-9: 0.009
100 C 5616-12: 0.027
5615-6: 0.028
5627A-12: 0.065
5610-9: 0.009
300 C 5616-12: 0.040
5615-6: 0.043
5627A-12: 0.103
5610-9: n/a
420 C 5616-12: 0.050
5615-6: n/a
5627A-12: 0.130
5610-9: n/a
Note 2: Includes readout accuracy, probe calibration, and probe drift
PRT Equivalent Temperature Accuracy (Readout only)
100 C 0.011
0 C 0.015
100 C 0.019
200 C   0.023
400 C 0.031
600 C 0.039
Thermistor Equivalent Temperature Accuracy (Readout Only)
0 C 0.002
25 C 0.003
50 C 0.006
75 C 0.014
100 C 0.030